Determination of phthalic acid esters in water samples by hollow fiber liquid-phase microextraction prior to gas chromatography tandem mass spectrometry
Fecha
2018Resumen
A new hollow fiber liquid-phase microextraction (HF-LPME) method has been developed for the
extraction of a group of phthalic acid esters (PAEs) of interest from different water samples prior to
gas chromatography tandem mass spectrometry analysis. HF-LPME was carried out using 1-octanol as
extraction solvent followed by a back extraction step with cyclohexane. The different parameters that
affect HF-LPME such as sample pH, ionic strength, extraction time, stirring rate, extraction temperature
and back extraction conditions were investigated. The optimized conditions involved the extraction of
10 mL of sample without pH adjustment or addition of salt during 75 min under a stirring of 850 rpm at
60 C and subsequent desorption with 200 mL of cyclohexane for 10 min in an ultrasonic bath. The
method was validated in terms of calibration and recovery studies using dibutyl phthalate-d4 as internal
standard. The developed procedure gave satisfactory recovery (74e120%) and relative standard deviation
values (<20%) for the studied PAEs in mineral, tap, pond and waste water samples